Internal Grade A laminar flow environment
-
Radiopharmaceutical dispensing requirements
-
Complies with GMP, radiation protection, and environmental standards
-
Diversified radiopharmaceutical research and development
Value
-
Air Velocity
Optimized inlet design ensures more stable airflow. The average downward laminar airflow velocity is maintained at 0.45 m/s ±10%, which exceeds the GMP 2010 Grade A requirement of 0.45 m/s ±20%.
-
Airflow Pattern
Optimized return air channel design provides a more uniform airflow pattern. Third-party professional airflow testing was conducted, including a static 6-point full-coverage test within a 1 m² chamber and a dynamic test at the operation position. All test points met the required standards.
-
Particle Control
Through optimized connector and air-tight sealing designs, along with a well-planned filter system layout, the internal chamber maintains a pressure of ≥ -20 Pa. Under this condition, particle counts were:
≥0.5 μm: 5 particles (/m³)
≥5.0 μm: 0 particles (/m³)
These results fully meet the GMP 2010 Grade A cleanliness requirements.
Customer Reviews
Q&A
-
What does the after-sales service for Mortenon H1 include?We provide comprehensive after-sales services for the Mortenon M1, including equipment installation and commissioning, operator training, regular maintenance, troubleshooting, and remote technical support. Additionally, spare parts are supplied promptly to ensure stable and worry-free operation. Our service team responds quickly, and users can contact us anytime via hotline, email, or other channels. We will arrange timely support based on the situation to ensure the equipment runs efficiently and reliably.